Mounting
2.4

Mount electronics of high-temperature sensors

The electronics of high-temperature sensors are attached to the end of a 32" (812 mm)
pre-installed flexible conduit. The electronics must be separately mounted on a wall or
instrument pole.
Figure 2-5:
A.
B.
C.
D.
With some large meter sizes, the meter may ship with the electronics attached to the
sensor case. The meter cannot be operated in this configuration. Detach the electronics
bracket from the sensor case and then proceed to mount the electronics to a wall or
instrument pole as described below.
Important
Do not operate the meter while the electronics are attached to the sensor case.
12
Components of a high-temperature sensor
A
B
C
D
Sensor
Electronics
Mounting bracket
Flexible conduit (minimum bend radius 6" [152 mm])
Micro Motion ELITE