2.3
Rotate junction box or 800 core processor
(optional)
An integrally mounted junction box or 800 core processor can be rotated to one of eight
possible positions in 45 degree increments.
Figure 2-4:
A.
B.
C.
D.
E.
Note
The 800 core processor is shown in this figure. The junction box has a somewhat different appearance.
Procedure
1.
2.
3.
4.
5.
Installation Manual
Parts for rotating the junction box or 800 core processor on the sensor
A
B
C
Housing
Clamping ring
Clamping ring screw
Feedthrough
Alignment notches
Loosen the clamping ring screw and remove the clamping ring.
Gently separate the housing from the feedthrough, but only until there is sufficient
clearance from the alignment notches to rotate the housing.
Rotate the housing to the desired position and in line with the alignment notches.
Seat the housing onto the feedthrough.
Replace the clamping ring and tighten the clamping ring screw.
D
Mounting
E
11