C A U T I O N
C A U T I O N
1 Apply pure isopropyl alcohol to a clean lint-free cotton swab or lens paper.
Hewlett-Packard strongly recommends that index matching compounds not be
applied to their instruments and accessories. Some compounds, such as gels,
may be difficult to remove and can contain damaging particulates. If you think
the use of such compounds is necessary, refer to the compound manufacturer
for information on application and cleaning procedures.
Table 4-2. Cleaning Accessories
Item
Isopropyl alcohol
Cotton swabs
Small foam swabs
Compressed dust remover (non-residue)
Table 4-3. Dust Caps Provided with Lightwave Instruments
Item
Laser shutter cap
FC/PC dust cap
Biconic dust cap
DIN dust cap
HMS10/HP dust cap
ST dust cap
To clean a non-lensed connector
Do not use any type of foam swab to clean optical fiber ends. Foam swabs can
leave filmy deposits on fiber ends that can degrade performance.
Cotton swabs can be used as long as no cotton fibers remain on the fiber end
after cleaning.
Cleaning Connections for Accurate Measurements
HP Part Number
8500-5344
8520-0023
9300-1223
8500-5262
HP Part Number
08145-64521
08154-44102
08154-44105
5040-9364
5040-9361
5040-9366
Reference
4-17