Overview

This chapter describes how to verify the performance of the Efficia DFM100 monitor/defibrillator after
repairs are complete. This chapter is organized into the following sections:

Required Testing Levels

Verification Test Equipment
Test and Inspection Matrix
Performance Verification Procedures
Required Testing Levels
The Performance Verification and Safety Tests are intended to verify proper operation of the device
following repair. The level of testing required corresponds to the type of repair performed.

External Repairs/Replacements

External repairs/replacements are those involving the repair or replacement of one or more of the
following items. The key point is that the case has not been opened.
Battery Compartment Cover and Latch
Bedrail Mount
Labels
Therapy Knob / Smart Select Knob
Paddle Tray and Plates
Handle
The following tests are required after an External Repair or Replacement when the case has not been
opened:
Perform the Visual Inspection (see
Run the Operational Check (see
Check the Error Logs for critical errors (see
After the Therapy Knob replacement only: Run the Controls test (see
After the Paddle Tray and Plates replacement only: Run the Paddle Safety Check (see
Check"
Performance Verification
.
.
.
.
on page 168).
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
"Visual Inspection"
on page 158).
"Operational Check"
on page 159).
"Error Log Messages"
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
on page 44).
"Controls Test"
5
.
. p. 151
.
. p. 152
.
. p. 153
.
. p. 157
on page 160).
"Paddles Safety
151