5: AED Mode Option
Table 15
AED Mode Shock Related Messages (Continued)
These messages may appear when a shock is advised or immediately after a shock is delivered in AED Mode.
Audio Message
"
Press pads firmly to patient's
"
bare skin.
None
None
Table 16
AED Mode Forced Pause Related Messages
These messages may appear during a forced pause time period when analysis is not taking place and you can attend
to the patient.
Audio Message
"
"
Begin CPR.
"
"
Attend to patient.
"
"
No shock advised.
"
Begin CPR, Check patient.
"
"
Stop CPR.

Using AED Mode to Monitor

You can use AED Mode to monitor your patient's ECG, SpO
alarms can also be activated for the parameters.
To monitor ECG in AED Mode and activate alarms:
1
Turn the Therapy knob to AED. The Efficia DFM100 announces the patient category currently
set.
If not correct, use the Patient Category button
74
Display Text
None




Display Text






"

None
Using AED Mode to Monitor
Condition
Shock delivered abnormal
energy due to marginal
impedance.
The final shock in a series
has been delivered.
A shock in a
yet-to-be-completed series
has been delivered.
Condition
A shock series has ended and
the device has entered a
pause period for CPR.
A no shock decision has
been made and the device
has entered a pause period
for CPR.
A no shock decision has
been made and during
monitoring, artifact has
been detected.
You have paused the device.
A no shock decision has
been made and the device
has entered a pause period
for CPR.
The CPR pause period has
ended.
, pulse, EtCO
2
to select the appropriate patient category.
User Action
Confirm pad
placement is correct
and pressed firmly
against the chest.
Attend to the patient.
Stay clear of patient.
User Action
Attend to the patient.
Begin CPR if
indicated.
Attend to patient.
Resume analysis if
indicated.
and AwRR. Related
2

Advertisement