4: ECG Monitoring
If needed, select the appropriate ECG wave size and press the Smart Select knob.
5
NOTE: When you select the ECG wave size for a certain lead, all instances of that lead adjust to the selected
size. For example, if you have Lead II selected for both Wave Sectors 1 and 2, and you change the
size in Wave Sector 1, Wave Sector 2's size changes automatically.
Figure 43

NOTES: Selecting the
clipping the wave sector.
Adjusting the ECG wave size on the display only affects the wave size on the display for viewing. It
does not affect the ECG signal used for arrhythmia analysis. Detected R-Waves for Synchronized
Cardioversion and Demand Pacing are also unaffected by the ECG wave size.
In AED Mode
The Pads ECG wave is automatically populated in Wave Sector 1 in AED Mode. The waves
configured for use in AED Mode are placed in Wave Sector 2 using the Smart Select knob.
To select a waveform for Wave Sector 2 in AED Mode:
1
Confirm your device is configured to monitor CO
2
From AED Mode, press the Smart Select knob.
Turn the Smart Select knob to highlight Wave 2 and press the Smart Select knob.
3
Select the wave you want placed in Wave Sector 2 and press the Smart Select knob.
4
Figure 44
54
Wave Menus






size, automatically adjusts the ECG size to the maximum size allowed without
Wave 2 Menu in AED Mode












and/or SpO
in AED Mode.
2
2




Monitor View





