Download  Print this page

Back Panel - Philips Efficia DFM100 Instructions For Use Manual

Defibrillator/monitor.
Hide thumbs

Advertisement

2: Device Basics

Back Panel

The back panel of the Efficia DFM100 has a compartment for the Lithium Ion battery. It also
contains the AC power connection, the ECG Out jack to connect to an external monitor, the USB
port and the LAN port. See
on page 16
Figure 15
LAN Port
USB Port
* - When the Efficia DFM100 is used together with other devices, their equalization terminals can be connected together to
eliminate any potential electrical differences between the two.
WARNING: Do not connect a LAN cable to the Efficia DFM100 while in a clinical mode. Incorrect ECG
diagnosis may result due to excessive electrical background noise.
Installing the Battery
To install the Efficia DFM100 Lithium Ion battery (see
1
Align the battery in the battery compartment. Confirm the arrow on the Battery Tab is pointed
up.
Push the battery into the battery compartment until you hear the Battery Latch lock into place.
2
Figure 16
14
Figure
15. For more information on ECG Out, see
.
Back Panel
Equalization
ECG Out Port
Terminal
Installing the Battery
AC Connection
*
Figure
16):
Basic Orientation
"ECG Out Cable"
Cable Strap
/Rear Pouch
connectors
Bedrail Hook
connection
AC Power
Cord Guard
Battery Latch
Battery
Compartment