19: Specifications
Table 63
Electromagnetic Immunity - General
Immunity Test
±
Electrostatic
discharge (ESD)
±
IEC 61000-4-2/
GB/T17626.2
Electrical fast
±
transient/burst
±
1 kV for input/output lines
IEC 61000-4-4/
GB/T17626.4
±
Surge
IEC 61000-4-5/
±
GB/T17626.5
Voltage dips, short
< 5% U
interruptions, and
(> 95% dip in U
voltage variations on
cycle
power supply input
40% U
lines
(60% dip in U
IEC 61000-4-11/
for 5 cycles
GB/T17626.11
70% U
(30% dip in U
for 25 cycles
< 5% U
(> 95% dip in U
Power frequency
3 A/m
(50/60 Hz) magnetic
field IEC 61000-4-8/
GB/T17626.8
U
is the AC mains voltage prior to application of the test level.
T
216
Test Level
6 kV contact
8 kV air
2 kV for power supply lines
1 kV differential mode
2 kV common mode
T
) for 0,5
T
T
)
T
T
)
T
T
) for 5 sec
T
Compliance Level
±
6 kV contact
±
8 kV air
±
2 kV for power supply lines
+1 kV for input/output lines
±
1 kV differential mode
+ 2 kV common mode
< 5% U
T
(> 95% dip in U
) for 0,5
T
cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
< 5% U
T
(> 95% dip in U
) for 5 sec
T
3 A/m
Electromagnetic Compatibility
Electromagnetic Environment
- Guidance
Floors should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital
environment.