Advertisement

Specifications

Thermal Array Printer

Continuous ECG Strip: The Print key starts and stops the strip. The printer can be configured to be
run real time or with a 10-second delay. The strip prints the primary ECG lead and a second wave
with event annotations and measurements.
Auto Printing: The printer can be configured to automatically print on Mark Events, Charge, Shock
and Alarm.
Reports: The following can be printed:
Event Summary (Long or Short)
Vital Signs Trends
Operational Check
Configuration
Status Log
Device Information
Speed: 25 mm/s with an accuracy of ±5%
Amplitude Accuracy: 5% for offset voltages of ± 300 mV at 5Hz
Paper Size: 50 mm (W) x 20 m (L)

Noninvasive Pacing

Waveform: Monophasic
Current Pulse Amplitude: 10 mA to 200 mA if the pulse width is set to 20 ms (5 mA increments);
accuracy ±10% or ±5 mA whichever is greater. For a 40 ms setting, the maximum pacing current is
140 mA.
Pulse Duration: 20 or 40 msec with ±10% accuracy
Rate: 30 ppm to 180 ppm (10 ppm increments); accuracy ±1.5%
Mode: Demand or Fixed
Refractory Period: 340 msec (30 to 80ppm); 240 msec (90 to 180 ppm) ±10%
Universal-function electrodes (Pads): After 60 minutes of pacing with approved defibrillators, the
Multifunction Electrodes (Pads) exhibit a post-defibrillation DC Offset of less than ± 800 mV at
seconds post-shock.
19: Specifications
4
205