19: Specifications

AED Mode

AED Energy Profile: 150 Joules (factory default) for Adult /50 J for Infant/Child nominal into a 50
ohm test load.
AED Controls: On/Off, shock.
Text and Voice Prompts: Extensive text/audible messages guide user through a user-configured
protocol.
Indicators: Monitor display messages and prompts, voice prompts, battery status, RFU, external
power.
Armed Indicators: Charging/charged tones, flashing shock button, energy level indicated on the
display.
ECG analysis: Evaluates patient ECG and signal quality to determine if a shock is appropriate and
evaluates connection impedance for proper defibrillation pad contact.
Shockable Rhythms: SMART Analysis is designed to shock ventricular fibrillation, ventricular
flutter and polymorphic ventricular tachycardia. It is designed to avoid delivering a shock for
rhythms that are commonly accompanied by a pulse or rhythms that would not benefit from an
electrical shock.
Shock Advisory Algorithm Sensitivity: Meets AAMI DF39 requirements and AHA
recommendations; Adult: Ventricular Fibrillation - 90% with lower confidence limit (LCL) of 87%,
Polymorphic Ventricular Tachycardia and Ventricular Flutter - 75% with LCL of 67%;
Infant/Child: Ventricular Fibrillation - 90% with LCL of 87%.
Shock Advisory Algorithm Specificity: Meets AAMI DF39 requirements and AHA
recommendations; Normal Sinus Rhythm - 99% with LCL of 97%; Asystole - 95% with LCL of
92%; Other non-shockable Rhythms - 95% with LCL of 88%.

ECG and Arrhythmia Monitoring

Inputs: Up to 3 ECG waves may be viewed on the display and up to 2 waves printed simultaneously.
Lead I, II or III is obtained through the 3-wire ECG cable and separate monitoring electrodes. With
a 5-Lead ECG cable, leads aVR, aVL, aVF and V can also be obtained. Pads ECG is obtained
through two multifunction electrode pads.
Lead Fault: Messages and dashed lines appear on the display if an electrode or lead becomes
disconnected.
Pad Fault: Dashed line appears on the display if a pad becomes disconnected.
Heart Rate Display: Digital readout on the display from 16 to 300 bpm (Adult Patient Category) or
16 to 350 bpm (Infant/Child), with an accuracy of ±10% or ±5 bpm whichever is greater.
Heart Rate/Arrhythmia Alarms: HR high/low, Asystole, VFIB/V-TACH, VTACH, Extreme Tachy,
Extreme Brady, PVC rate, Pacer Not Capture, Pacer Not Pacing.
Common Mode Rejection: 105 dB for Leads ECG, 96 dB for pads ECG.
ECG Size: 1/4x, 1/2x, 1x, 2x, 4x, auto gain (1x gain is 10mm/mV on the printed strip).
ECG waveforms: Displayed at a fixed timebase of 25 mm/sec (printer) ±5%, 25 mm/sec (display)
±10%.
ECG Leads Off Sensing: 3- and 5-Lead wires apply a <35nA DC current patient electrodes,
<1.0uA other electrodes.