19: Specifications
Table 58
Selected
Energy
1 J
2 J
3 J
4 J
5 J
6 J
7 J
8 J
9 J
10 J
15 J
20 J
30 J
50 J
70 J
100 J
120 J
150 J
170 J
200 J
The delivered energy accuracy is ±10% or ±1J whichever is greater for all energy settings.
Charge times:
Less than 5 seconds to the recommended adult energy level (150 Joules) with a new fully-charged
battery installed.
Less than 6 seconds to the selected energy level (up to 200 Joules) with a new fully-charged battery
installed, even after the delivery of 15 discharges at maximum energy.
Less than 15 seconds to the selected energy level while connected to AC power only, even when
operating on 90% of the rated mains voltage.
The device powers on in manual defibrillation mode ready to deliver shock in less than:
200
Delivered Energy Accuracy
Nominal Delivered Energy vs. Load Impedance
Load Impedance (ohms) ±2%
25
50
1.2
1.3
1.3
1.7
2.0
2.1
2.6
3.0
3.1
3.5
4.0
4.2
4.3
5.0
5.2
5.2
6.0
6.3
6.1
7.0
7.3
6.9
8.0
8.4
7.8
9.0
9.4
8.7
10
10
13
15
16
17
20
21
26
30
31
43
50
52
61
70
73
87
100
105
104
120
126
130
150
157
147
170
178
173
200
209
23 seconds with AC power only and at 90% of rated mains voltage.
15 seconds with a new, fully-charged battery even after 15 discharges of maximum energy.
75
100
125
150
1.2
1.1
1.0
2.0
1.9
1.7
3.2
3.2
3.1
4.3
4.4
4.5
5.4
5.5
5.6
6.5
6.6
6.7
7.6
7.8
7.8
8.6
8.9
8.9
9.7
10
10
11
11
11
16
17
17
22
22
22
32
33
33
54
55
56
76
78
78
108
111
111
130
133
134
162
166
167
184
188
189
216
222
223
Specifications
175
0.9
1.6
2.9
4.3
5.4
6.5
7.6
8.7
9.8
11
16
22
33
54
76
108
130
163
184
217

Advertisement

Table of Contents

   Also See for Philips Efficia DFM100

Comments to this Manuals

Symbols: 0
Latest comments: