17: Troubleshooting
Table 54
NBP Problems (Continued)
Symptom

technical alarm.

technical alarm.
Table 55
Printing Problems
Symptom
Paper won't move.
Paper moves and then stops.
Paper moves but printing is
faint or absent.
Paper moves but print
quality is poor or some dots
are missing.
White line running along
paper.
Loud buzzing or grinding
noise.

technical
alarm.

technical
alarm.

technical alarm.

technical
alarm.

technical alarm. A non-critical device failure
192
Possible Cause
Faulty NBP hardware.
A non-critical device failure
has occurred.
Possible Cause
Paper improperly loaded,
jammed, or wet.
Door improperly latched.
Paper improperly loaded or
jammed.
Paper roll improperly
installed.
Incorrect paper type.
Printhead temperature
approaching maximum
recommended operating
temperature.
Dirty printhead.
Print door improperly
latched.
The printer has run out of
paper.
The printer door is not fully
closed.
The required font is
unavailable for the currently
installed language.
The printer is faulty or there
is a problem communicating
with the printer.
has occurred.
Potential Solution
Call for service. If NBP monitoring is essential to
patient care, take the device out of use.
Potential Solution
Reload paper or clear jam. If paper is wet, replace with a
fresh, dry roll.
Check door latch.
Reload paper or clear jam.
Check that the paper is installed correctly.
Use only recommended paper type.
Wait until the printer cools down to restart printing. If
there is a lot of black printed on the paper, check ECG
for excessive noise.
Clean the printhead.
Check door latch.
Reload with new fresh, dry roll of paper.
Open the printer door and reclose such that it snaps
into place.
If printing is essential to patient care, take the device out
of use and call for service.
Turn the Efficia DFM100 off for 15 seconds and then
turn it back on. If the problem persists, call for service.
If printing is essential to patient care, take the device out
of use.
Restart device. If the error continues and printing is
essential to patient care, take the device out of use and
call for service.