Philips Efficia DFM100 Instructions For Use Manual Page 203

Defibrillator/monitor.
Hide thumbs

Advertisement

Symptoms
Table 53
EtCO
Problems (Continued)
2
Symptom

technical alarm.

technical alarm.

technical alarm.

technical alarm.
Table 54
NBP Problems
Symptom
Measurement cycle doesn't
automatically start.
The pump operates but the
cuff does not inflate or fails
to inflate fully.
NBP measurements appear
high/low.

technical alarm.
The NBP numeric value is

replaced with a
.

technical alarm.
The NBP numeric value is

replaced with a
.

technical alarm.
The NBP numeric value is

replaced with a
.
Possible Cause
For Mainstream sensors -
your airway adapter is
blocked or not mounted
properly.
The CO
sampling line is
2
disconnected.
The CO
sensor is
2
unplugged.
A non-critical error was
detected.
Possible Cause
NBP is not configured for
automatic measurements.
Automatic measurements are
not scheduled for the current
patient.

The
soft key has
not been pressed.
Defective cuff.
Poor connection between the
cuff and the Efficia
DFM100.
The cuff size is too
small/large for the patient.
The cuff has not fully
deflated after 3 minutes.
The NBP cuff pressure has
exceeded the overpressure
safety limit of 300
mmHg/40 kPa.
A measurement value could
not be obtained.
Potential Solution
Clean airway adapter if mucus or moisture is seen.
Mount sensor correctly on the adapter. Zero the sensor.
Confirm the sampling line is firmly connected to the
CO
sensor.
2
Confirm that the CO
sensor is solidly connected to the
2
Efficia DFM100. If the sensor is solidly connected to
the Efficia DFM100 and the technical alarm remains on
the display, try to re-zero the sensor to clear the alarm.
Restart device when it is appropriate to do so. If the
alarm continues, replace the CO
Potential Solution
Check/modify the configuration as needed.
Use the Measurements/Alarms menu to define an
automatic schedule of measurements for the current
patient.

Press the
soft key.
Replace the cuff.
Check connections and replace tubing if needed.
Use the correct cuff size and take another measurement.
Remove the cuff from the patient.
Release pressure in the cuff (disconnect cuff from
tubing).
Replace the cuff. If the problem persists, call for service.
The cuff should deflate automatically. If not, remove
cuff from patient and deflate. Turn device off to reset
the alarm and restart NBP.
Check cuff size and placement.
17: Troubleshooting
sensor.
2
191