16: Maintenance

Printer Printhead

If the printout has light or varying print density, clean the printhead to remove any buildup of paper
residue.
To clean the printhead:
Open the printer door.
1
Move the paper out of the way.
2
3
Clean the printhead surface (top, front of the compartment) with a cotton swap dipped in
isopropyl alcohol.
Replace roll of paper and close door.
4

Side Pouches

After removing from the device, the side pouches may be cleaned by hand with mild soap and water
and air dried. Do not wash or dry by machine.
SpO
Sensor and Cable
2
Follow the manufacturer's instructions to clean the SpO
CO
Sensor and Cable
2
Follow the manufacturer's instructions to clean the CO

NBP Cuff

Follow the manufacturer's instructions to clean the cuff.

Service Life

Life expectancies for the Efficia DFM100 therapy cable and external paddles are related to the
environment they are used in, their frequency of use and how they are cared for. Their service life is
up to three years. To maintain reliable performance and reduce the possibility of failure during
patient use, replace them every three years from the time they were initially placed into service or if
they fail inspection. Extra care must be taken when using the Efficia DFM100 in transport and
Emergency Medical Service (EMS) environments that place additional stress on the cable
connection and increase the chance for wear, impacting its useful life.

Efficia DFM100 Disposal

Prior to disposal, remove the battery. Then dispose the device and accessories in accordance with
your country's regulations for equipment containing electronic parts.
WARNINGS: Disposing the device with the battery inserted presents an potential shock hazard.
To avoid contaminating or infecting personnel, the environment or other equipment, make sure you
disinfect and decontaminate the device and any appropriate accessories prior to disposal.