16: Maintenance

Power-Related Alarms

Power-related alarms are generated for the conditions shown in
appear as alarm messages on the Efficia DFM100 display. There are both audio and visual alerts. For
more information on alarms, see
Table 48
Alarm Message








176
"Alarms"
Power-Related Alarms
Condition
Battery power is low.
Battery power is critically low.
The device will shut down in
1 minute.
Battery power is critically low.
The device is shutting down
now.
A low voltage has been
detected.
Communications between the
device and the battery have
failed.
The battery has reached its
end of life.
Table
on page 36.
Type of Alarm
High Priority if
pacing otherwise,
low priority
non-latching alarm
High Priority if
pacing otherwise,
medium priority
non-latching alarm
High Priority
Non-Latching
Medium Priority
Non-Latching
Low Priority
Non-Latching
Battery Maintenance
48. Once generated, they
Indication and
Location
Red alarm message if
pacing, cyan if not
with audio tone in
Battery Status area.
Red alarm message if
pacing, yellow if not
with audio tone in
Technical Alarm
area.
Red alarm message
with audio tone in
Technical Alarm
area.
Yellow alarm message
with audio tone in
Battery Status area.
Cyan alarm message
with audio tone in
Technical Alarm
area.