Operational Check
During the Operational Check, when a response is required, use the Smart Select knob to select
9
your answer and then confirm your choice. As each test runs, the name of the test appears
highlighted on the display with the message
Figure 75
NOTE: Once you have pressed the Sync button, you can leave the Efficia DFM100 unattended as
Operational Check completes its process. If you cancel the Operational Check before it completes all
tasks, there is no record of the check in the Operational Check Summary.
After the automated part of Operational Check concludes, an Operational Check Report is printed
(see
"Printing Operational Check Results"
NOTE: It is important to complete all instructions listed on the Setup screen in order to successfully
complete an Operational Check. Approximately 10 seconds after entering Operational Check, a

continue with the Operational Check if the device is not responding to actions you have taken
during Setup. Selecting
instructions causes Operational Check failures.
Operational Check In Process Screen


 









prompt appears in the lower right corner of the display. This prompt allows you to

15: Operational and Shift Check

. See


















on page 166).
while there are still required items listed in the setup
Figure
75.
