Philips Efficia DFM100 Instructions For Use Manual Page 164

Defibrillator/monitor.
Hide thumbs

Advertisement

14: Configuration
Table 34
NBP Settings (Continued)
Parameter

Selects the high limit alarm value when systolic is

the selected alarm source.

Selects the low limit alarm value when systolic is

the selected alarm source.

Selects the high limit alarm value when diastolic

is the selected alarm source.

Selects the low limit alarm value when diastolic is

the selected alarm source.

Selects the high limit alarm value when mean is
the selected alarm source.

Selects the low limit alarm value when mean is
the selected alarm source.
152
Description
Setting Choices
Adult: 35-255, 160 (mmHg);
4.5-34, 21 (kPa)
Infant/Child: 35-135, 120
(mmHg); 4.5-18, 16 (kPa)
adjusted in increments of 5
mmHg/0.5 kPa
Adult: 30-250, 90 (mmHg);
4-33.5, 12 (kPa)
Infant/Child: 30-130, 70
(mmHg); 4-17.5, 9 (kPa)
adjusted in increments of 5
mmHg/0.5 kPa
Adult: 15-220, 90 (mmHg);
2-29.5, 12 (kPa)
Infant/Child: 15-110, 70
(mmHg); 2-15, 9 (kPa)
adjusted in increments of 5
mmHg/0.5 kPa
Adult: 10-215, 50 (mmHg);
1.5-29, 7 (kPa)
Infant/Child: 10-105, 40
(mmHg); 1.5-14.5, 5 (kPa)
adjusted in increments of 5
mmHg/0.5 kPa
Adult: 25-235, 110 (mmHg);
3.5-31.5, 15 (kPa)
Infant/Child: 25-125, 90
(mmHg); 3.5-16.5, 12 (kPa)
adjusted in increments of 5
mmHg/0.5 kPa
Adult: 20-230, 60 (mmHg);
3-31, 8 (kPa)
Infant/Child: 20-120, 50
(mmHg); 3-16, 7 (kPa)
adjusted in increments of 5
mmHg/0.5 kPa
Overview
User Setting