13: Data Management
Table 29
Event Information (Continued)
Logged Event

(low value,
high value)

(low value,
high value)

(low value,
high value)

(value, limit)

(value, limit)

(value, limit)

(value, limit)

(value, limit)

(value, limit)





Defibrillation

(value)


(reason)

(number) (energy)
(impedance) (peak current)

(impedance)


/

136
Logged when the Systolic NBP alarm limit is changed.
Logged when the Diastolic NBP alarm limit is changed.
Logged when the Mean NBP alarm limit is changed.
Logged when the patient's Systolic NBP is higher than the configured limit.
Logged when the patient's Systolic NBP is lower than the configured limit.
Logged when the patient's Diastolic NBP is higher than the configured limit.
Logged when the patient's Diastolic NBP is lower than the configured limit.
Logged when the patient's Mean NBP is higher than the configured limit.
Logged when the patient's Mean NBP is lower than the configured limit.
Logged when the NBP cuff fails to deflate after 3 minutes.
Logged when the NBP cuff pressure reaches 300 mmHg/40 kPa.
Logged when an NBP measurement fails to complete.
Logged when the device detects a non-critical NBP failure.
Logged when the NBP module detects a malfunction.
Logged when charging is initiated.
Logged when the device is disarmed manually.
Logged when the device disarms automatically. Reasons include:

: Logged when the automatic disarm is caused by a bad connection between
the device and the patient.

energy during charging.

: Logged when the device reaches its configured auto-disarm period.

: Logged when, in AED Mode, the algorithm determines the
rhythm is not shockable.

: Logged when the device is turned off while charged.

: Logged in Synchronized Cardioversion Mode when a leads off condition is
detected in the synchronizing lead.

device detects a change in the paddles or pads type or if the therapy cable type
identification is not valid.
Logged when a shock is delivered.
Logged when a shock is initiated but aborted before a full shock dose is delivered.
Logged when Sync is turned on or off and any subsequent change.
Logged when a shock is initiated and completed but the full dose is not delivered.
Frequency
: Logged when the device is unable to reach the selected
: Logged when, with the therapy cable connected, the
Event Summary