This chapter describes how to review patient data using Efficia DFM100 Trending.

Overview

In Monitor Mode your Efficia DFM100 provides the ability to view and print numeric vital signs
trending data for the current incident. Trending data are automatically acquired if parameters are on.
When viewing trending data, the Trending Report is displayed in the Efficia DFM100's lower two
wave sectors and takes over the soft key functions. Trend data can be displayed at selected intervals
for up to 8 hours of monitoring. You can set your trend interval to 1, 5, 10, 15, 30 or 60 minutes.
Trending data displayed for parameters continuously measured (heart rate, SpO
are the average of multiple measurements over the trend time period. Trending data for NBP appear
with the timestamp of the measurement.

Viewing Trend Data

To view trending data:
Confirm your Efficia DFM100 is in Monitor Mode.
1
2
Press the Smart Select knob.
3
Turn the Smart Select knob to highlight
takes over the bottom two wave sectors. See
To close the Trends Report, press the
Figure 64

Trending Report




























and press the knob in. The Trending Report
Figure
64.

soft key.

     

























Trending
, EtCO
2












12
and pulse)
2
125