11: Monitoring NBP

NBP Alarms

Alarms are annunciated when a measurement for the configured source (systolic, diastolic or mean)
falls outside the configured high or low limits. NBP alarms are non-latching alarms, meaning they
are automatically removed when their alarm condition no longer exists. Both the source of the alarm
and the limits may be changed during an ongoing patient event.
NBP alarms can be generated for the conditions shown in Tables
appear as alarm messages in the NBP alarm status area above the NBP numeric. There are both
audio and visual alerts. For more information on alarms, see
NOTE: Alarm notification is configurable. See
Table 27
Alarm Message






Table 28
Alarm Message









122
NBP Physiological Alarms
Condition
The NBP systolic value exceeds
the high alarm limit.
The NBP diastolic value exceeds
the high alarm limit.
The NBP mean value exceeds the
high alarm limit.
The NBP systolic value has fallen
below the low alarm limit.
The NBP diastolic value has
fallen below the low alarm limit.
The NBP mean value has fallen
below the low alarm limit.
NBP Technical Alarms
Condition
The cuff pressure exceeded 300
mmHg/40 kPa.
The cuff has failed to deflate.
The device is unable to complete
a measurement.
A problem with the NBP
module has been detected.
A non-critical failure has been
detected.
"Alarms"
"Alarm Management"
on page 38.
Type of Alarm
Medium
Priority
Latching
Configurable
Alarm
Type of Alarm
Low Priority
Non-Latching
Alarm
NBP Alarms
27
and 28. Once generated, they
on page 36.
Indication
Yellow alarm message
with audio tone
Indication
Cyan alarm message with
audio tone