Changing The Apnea Time Alarm Limit; Enabling/disabling Awrr Alarms; Zeroing Sidestream And Mainstream Sensors - Philips Efficia DFM100 Instructions For Use Manual

Defibrillator/monitor.
Hide thumbs

Advertisement

9: Monitoring CO
2

Changing the Apnea Time Alarm Limit

To change the apnea time alarm limit:
Press the Smart Select knob.
1
2
Turn the Smart Select knob to highlight the
Select knob.
Select
3
Select
4
Turn the Smart Select knob to select the new high limit value then press the Smart Select knob.
5
Set the new low limit value and press the Smart Select knob.
6

Enabling/Disabling AwRR Alarms

Press the Smart Select knob.
1
Turn the Smart Select knob to highlight the
2
Select knob.
Select
3
4
Select
WARNING: The safety and effectiveness of the respiration measurement method in the detection of apnea,
particularly the apnea of infancy, has not been established.

Zeroing Sidestream and Mainstream Sensors

To avoid inaccurate readings, Sidestream and Mainstream sensors need to be reset and require a valid
zero be performed when:
• attaching a new sample line
• there has been a significant change in environmental conditions
• when the accuracy of the reading is questionable
• requested by the Efficia DFM100
NOTES: Do not zero Sidestream and Mainstream CO
Wait 20 seconds after removing the sampling line from the patient's airway before zeroing the CO
sensor, so any lingering CO
Keep the sampling line away from all sources of CO
patient's) and ventilator exhaust.
During zeroing, EtCO
appears in the wave sector.
106

and press the Smart Select knob.

and press the Smart Select knob.

and press the Smart Select knob.


(
) and press the Smart Select knob.
in the line can dissipate.
2
data is invalid. A
2
Zeroing Sidestream and Mainstream Sensors


sensors without a sampling line installed.
2
, including exhaled breaths (yours and the
2

is displayed in the parameter block and a dashed line
menu and press the Smart
menu and press the Smart
2