8: Pacing

Pacing View

Pacing View appears when the Therapy knob is turned to the Pacer position. Pacing View includes a
status block which appears in Wave Sector 3 of the display (see
paced status is not displayed. Pacing-related information in Pacing View includes:
Pacing Markers: Markers, indicating a pace pulse was delivered, appear in Wave Sector 1 (and
in Wave Sector 2, if the wave is cascading) each time a pacer pulse is delivered.
R-Wave Arrows: R-Wave arrows appear in Wave Sector 1 (and in Wave Sector 2, if the wave is
cascading) when in Demand Mode pacing. R-Wave arrows do not appear on paced beats.
Pacing Status: Indicates the current pacing status.
Pacing Alarm: If there is a pacing-related alarm during pacing, the current pacing status is
replaced with an alarm message. See
Pacing Mode: Indicates if the device is in Demand or Fixed Mode Pacing.
Pacing Rate: Indicates the current pacing rate, including unit of measure.
Pacing Output: Indicates the current output, including unit of measure.
Alarms: Are on automatically.
Figure 53
Pacing View Layout






Pacing
Markers


Pacing Bar:
Mode,

Status,
Alarms


Pacing Control soft key: Press
to pause pacing; press
resume pacing
WARNING: If pacing is interrupted for any reason, you must press the
92
When pacing is active,
If the device is running on battery,
If pacing is not active,













to start or

is displayed when the device is on AC power.


is displayed.
"Pacing Alarms"

















Pacing
Rate Value
Pacer Settings soft key: Press
bring up the menu for adjusting Pacer Rate and
Output.
Figure
53). The patient's internal
is displayed.
on page 97.



 









Pacing
Output Value


soft key to resume pacing.
Pacing View

R-Wave
Arrows
to