Download Print this page

Aseptico CompriCart ADU-20B Operation And Maintenance Instruction Manual

Hide thumbs Also See for CompriCart ADU-20B:

Advertisement

OPERATION and MAINTENANCE
INSTRUCTION MANUAL
ADU-20B CompriCart

Advertisement

loading
Need help?

Need help?

Do you have a question about the CompriCart ADU-20B and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Aseptico CompriCart ADU-20B

  • Page 1 OPERATION and MAINTENANCE INSTRUCTION MANUAL ADU-20B CompriCart...
  • Page 2: Table Of Contents

    International (425) 487-3157 • Toll Free(800) 426-5913 - Fig. 3 - HVE Vacuum Assembly ..3 www.aseptico.com • info@aseptico.com - Fig. 4 - Saliva Ejector Assembly ..3 - Fig. 5 - Water Supply Tank ...3 To prevent injury to people and damage to Operation Functions .
  • Page 3: Introduction

    Your new Aseptico ADU-20B CompriCart delivery system is one of the finest portable carts available to the dental profession. The system is designed for portable operatory use and features a three-way air/water syringe, automatic highspeed and lowspeed handpiece controls, high volume evacuator and saliva ejector.
  • Page 4: Setting Up The Unit

    SETTING UP THE SYSTEM: 1. Place the ADU-20B CompriCart in an 3. The ADU-20B unit must be connected to a upright position on a flat, stable surface. dry, filtered, 70-100 psi air source. Simply Loosen the center column retaining knob extend the 1/4"...
  • Page 5 4. Remove the high volume vacuum system, Locking ring will snap into place. Press saliva ejector system, and High-Density release lever on disconnect to remove Polyethylene (HDPE) water supply tank from bottle from unit (Fig. 4b). packing and install on proper quick Figure 4a Figure 4b disconnects on the delivery head (Fig.
  • Page 6: Operation Functions

    APPLY PRESSURE TO 2. HANDPIECE CONTROLS (Fig. 7) - The ACTIVATE Aseptico ADU-20B includes two HANDPIECE handpieces mounted to the front bar. These are fully automatic. By placing a handpiece into its proper holder, the Figure 8 - Foot Control...
  • Page 7 4. WATER SUPPLY TANK - The Aseptico 6. WATER COOLANT FLOW CONTROL ADU-20B incorporates a self contained (Fig. 10) - Adjusts the flow of water coolant pressurized water system (Fig.
  • Page 8 OPERATION FUNCTIONS Cont’d HANDPIECE WATER COOLANT PRESSURE FLOW CONTROL CONNECTION GAUGE AIR SUPPLY FLUSH LINE TOGGLE AIR SUPPLY SALIVA EJECTOR PRESSURE OPTIONAL ON/OFF REGULATOR CAVITRON CONNECTION WATER SUPPLY TANK Figure 10 - Right Side of Chassis Waste from the vacuum system will collect QUICK in the attached plastic bottle.
  • Page 9 To operate, remove the saliva ejector valve clean water or vacuum system cleanser from the holder and insert a standard through the evacuator hose and thoroughly saliva ejector tip. Turn the vacuum toggle cleaning the waste bottle. switch, located on the right side of the cart, 10.
  • Page 10: Routine Adjustments

    ROUTINE ADJUSTMENTS: 1. WATER COOLANT FLOW ADJUSTMENT 3. HANDPIECE PRESSURE ADJUSTMENT The water coolant to the handpiece control (Fig. 15) - The amount of air pressure can be varied from a fine fog spray to a supplied to each handpiece may be heavy stream.
  • Page 11: Maintenance

    MAINTENANCE: Because of its simple design, the Aseptico 4. PACKING THE UNIT FOR ADU-20B CompriCart requires very little TRANSPORTATION - If the ADU-20B unit maintenance. Any maintenance that is is going to be transported, it is needed can be performed in minutes.
  • Page 12: Warranty

    Aseptico's sole obligation under product warranty is (at its sole option and discretion) to repair or replace any defective component or product in part or whole. Aseptico shall be the sole arbiter of such action.