Advertisement

5.4

Potting

The sensor can be potted with silicon dielectric gel (non-toxic, transparent and self-healing
gel)
NOTICE
Electrical connections
Do not pot meter before electrical connections have been made.
● Mix the two components of the potting kit well and pour into terminal box.
● Let cure for approximately 24 hours at approximately 25°C (77°F). Curing time increases
by 100% per -10°C (-18°F).
Horizontal orientation
Note
Gel can be penetrated with test instruments or be removed in case of cable replacement.
SITRANS F M MAG 1100 sensor
Operating Instructions, 01/2018, A5E03433301-AC
Vertical orientation
Connecting
5.4 Potting