Table 3-4. Complete Calibration Summary (1 of 3)
Calibration
What is calibrated
Mainframe Calibration
Accuracy and
continuity of the
timescale
O/E Factory
The photodetector
Wavelength
responsivity
Calibration
O/E User Wavelength
The photodetector
Calibration
responsivity
Plug-in Vertical
Vertical offset and
Calibration
vertical scale accuracy
for both electrical and
optical channels.

Complete Calibration Summary

Measurements
Affected
Channels affected:
optical & electrical. All
time base
measurements such
as rise time, fall time,
eye width, and jitter.
Channels affected:
optical. Amplitude
accuracy of all optical
channel
measurements.
Optical power meter
accuracy.
Channels affected:
optical. All optical
channel
measurements at user
wavelengths.
Channels affected:
optical & electrical.
Any optical or
electrical vertical
measurements such
as V p to p , eye height,
extinction ratio, and
the optical power
meter
Complete Calibration Summary
Recommended
Key Path
Interval
Annually at HP service
Utility
center or if operating
Calibrate
temp has changed and
Calibrate frame
remains 5°C or more
from calibration
temperature. See
service manual.
Annual factory re-
Not user accessible.
calibration of standard
wavelengths.
Annual re-calibration
Optical Channel Setup
of user defined non-
Calibrate
factory wavelengths
O/E Cal
Perform after any
Utility
power cycle or once
Calibrate
every 10 hours during
Calibrate Plug-in
continuous use or if
operating temperature
changes by more than
2°C.
Calibration Overview
a
3-19