Download  Print this page

Philips FM20/30 Instructions For Use Manual Page 94

Release j.3 with software revision j.3x.xx patient monitoring.
Hide thumbs

Advertisement

4 FM20/30 Battery Option
check the specific cause of the problem by looking at the symbol(s) displayed in the
window.
Battery Status Indicator
Battery Status Window
1
To access the
information on the screen, or select
Battery status/malfunction symbols: If a problem is detected with the battery, an INOP may be
issued, and the following symbols are displayed in the
accompanied by a status message providing more details.
94
Alternates with the battery gauge
on the main screen.
Check in the
Battery Status
window to see which status
symbol is displayed to identify
the cause.
window and its associated pop-up keys, select the battery status
Battery Status
,
tells you how much power is left in the battery.
Capacity
remaining
,
tells you how much power the battery can hold when fully charged.
Capacity
fullCharge
tells you approximately how long you can continue to use the monitor without
TimeToEmpty
an AC connection. The time span fluctuates depending on the system load (the display
brightness and how many measurements you carry out), and the remaining capacity of the
battery.
is shown in place of
TimeToFull
and tells you how much time is left until the battery is charged to 90%. If >10 hr is shown
here, the battery may not charge completely when the monitor is in use.
Battery Malfunction Indicator
,
.
Main Setup
Battery
if the monitor is connected to a power supply,
TimeToEmpty
Battery Status
Battery Status
The red exclamation mark
flashes. Critical battery situation
or malfunction. Check in the
window to see
Battery Status
which malfunction indicator is
displayed, or refer to the INOP,
to identify the cause.
window, where they may be