Disconnecting From Power; Power On/power Off Behavior; Monitoring After A Power Failure - Philips FM20/30 Instructions For Use Manual

Release j.3 with software revision j.3x.xx patient monitoring.
Hide thumbs

Advertisement

Disconnecting from Power

FM20/30
To disconnect the monitor from AC power, switch the monitor off using the On/Off switch located
on the right side of the device, or unplug the power cord from the AC mains socket.
FM20/FM30 with Battery Option and FM40/FM50
The On/Standby button does not disconnect the monitor from the AC power source. To disconnect,
unplug the power cord from the AC mains socket. Note that if the power cord is unplugged from the
AC mains socket before the monitor is put into Standby, a beeper is activated. The beeper warns you if
the monitor is accidentally disconnected from AC mains.

Power On/Power Off Behavior

The general rules determining the behavior of the fetal monitors when connected to, or disconnected
from power are as follows:
A fetal monitor that was switched on prior to a temporary power loss, switches on again when
power is restored.
A fetal monitor that was switched off prior to a temporary power loss, remains off when power is
restored.
When AC mains power is lost, a battery powered monitor (FM20/30) continues to run without
interruption on battery power.

Monitoring After a Power Failure

A fetal monitor that was switched on prior to a temporary power loss switches on again when
power is restored.
A fetal monitor that was switched off prior to a temporary power loss remains off when power is
restored.
If the fetal monitor is without power for less than one minute, monitoring will resume with all
active settings unchanged.
If the fetal monitor is without power for more than one minute, the behavior depends on your
configuration.
If
is set to
Automat. Default
If
is set to
Automat. Default
48 hours. The
Automat. Default
, the default profile will be loaded when power is restored.
Yes
, all active settings are retained, if power is restored within
No
setting is made in Configuration Mode.