3 Basic Operation
The trace records maternal parameters also. When measuring noninvasive blood pressure, the
annotation is made at the end of the measurement. If the noninvasive blood pressure measurement
repetition time is short, the noninvasive blood pressure numeric may not always be printed.
The recording of notes (see "Entering Notes" on page 77) or time/date information may be
interrupted by connecting, or unplugging a transducer, or by a change in measurement-related setting
(for example, artifact suppression, Toco sensitivity, or alarm settings).
A new patient admission or a change to the paper scale setting stops all annotations, and prompts a
new vertical trace header to be printed.
To switch off the recorder:
Either select
1
Or press one of the SmartKeys (depending on configuration): fetal recorder
2
Recordng
If your recorder is configured with
need to confirm that you want to stop the recorder, before it will stop.
When the recorder is off, the "recorder off" status indicator is displayed in the bottom right-hand
corner of the screen: When the
and when
66
from the
Start/Stop
Fetal Recorder
.
Paper Save Mode
is
the icon shows a paper trace icon.
Paper Save Mode
On
1
menu.
on (a Configuration Mode setting), you will
Confirmed Stop
is set to
Off
Time stamp printed every ten
minutes
Start/ Stop
the paper symbol shows an x mark,
or
Stop