29 Default Settings Appendix
NBP Default Settings
Factory Default Settings
Mode
Repeat Time
Alarms from
Low Limit
High Limit
VP Pressure
Done Tone
Veni Puncture
Start Time
NBP
Alarms
Color
Reference
CL NBP Default Settings
NBP Setting
Mode
Repeat Time
NBP
VP Pressure
Reference
Unit
Done Tone
Start Time
Aging Time
Color
296
Manual
15 min
Sys.
90/50 (60)
160/90 (110)
60 mmHg
Off
n/a
Synchronized
On
On
Red
Auscultatory
Factory Defaults
Auto
10 min
On
60 mmHg
Auscultatory
mmHg
Off
Synchronized
10 min
Red