Displayed Temperature Measurement Range
Mode
Rectal (ear + 1.16°C)
Caution: ASTM E1965-98 specifies 34.4°C—42.2°C (94°F—108°F)
Ambient Temperature Range
Mode
Operating 10%—95% RH, non-
condensing
Storage up to 95% RH, non-
condensing
Caution: EN ISO 80601-2-56 specifies 16.0°C—35.0°C (60.8°F—95.0°F), 10%—95% RH, non-condensing
ASTM E1965-98 specifies 16.0°C—40.0°C (60.8°F—104.0°F), up to 95% RH, non-condensing
Storing the thermometer outside the specified temperature/humidity range might adversely affect measurement
accuracy. Check the calibration after storage in uncertain conditions.
Tympanic Temperature Alarm Specifications
Range
Adjustment
Alarm delay
Physical Specifications
Thermometer
Dimensions
Cable length
Weight (including cable)
Ingress protection classification
Base Station
Dimensions
Weight (excluding cable)
Range °C
34.2—42.0
Range °C
16.0—33.0
-25.0—55.0
33.0°C—42.0°C (91.0°F—108.0°F)
0.5°C steps (33.0°C—35.0°C)
0.1°C steps (35.0°C—42.0°C)
1.0°F steps (91.0°F—95.0°F)
0.2°F steps (95.0°F—108.0°F)
System alarm delay (see "Fetal Monitors" on page 267).
190 mm x 43 mm x 55 mm (±3 mm)
60 ±5 cm (spiral cable relaxed)
250 ±15 cm (spiral cable extended)
180 ±10 g
IP 21
205 mm x 65 mm x 75 mm (±3 mm)
400 g ±10 g
28 Specifications and Standards Compliance
Range °F
93.6—107.6
Range °F
60.8—91.4
-13.0—131.0
277