Philips FM20/30 Instructions For Use Manual Page 273

Release j.3 with software revision j.3x.xx patient monitoring.
Hide thumbs

Advertisement

Maternal ECG Supplemental Information as required by IEC 60601-2-27
Response Time of Heart Rate
Meter to Change in Heart Rate
Tall T-Wave Rejection Capability
Response Time of Heart Rate
meter to Change in Heart Rate
Heart Rate Meter Accuracy and
Response to Irregular Rhythm
Noninvasive Blood Pressure
Complies with IEC 80601-2-30:2009 / EN 80601-2-30:2010.
Performance Specifications
Measurement
Ranges
1
Accuracy
Pulse Rate
Measurement Time
Cuff Inflation Time
Initial Cuff Inflation Pressure
HR change from 80—120 bpm: 10 seconds
HR change from 80—40 bpm: 14 seconds
M2735A
M2738A
+
CL Toco
MP,
CL ECG/IUP
M2735A, M2738A HR change from 80—120 bpm Average: 12 seconds
+
CL Toco
MP,
CL ECG/IUP
M2735A, M2738A Ventricular bigeminy 40—60 bpm
+
CL Toco
MP,
CL ECG/IUP
Systolic
Diastolic
Mean
Range
Accuracy
(average over
noninvasive blood
pressure measurement
cycle)
28 Specifications and Standards Compliance
1.2 mV T-Wave amplitude
1.4 mV T-Wave amplitude
1.2 mV T-Wave amplitude
HR change from 80—40 bpm Average: 15 seconds
HR change from 80—120 bpm Average: 10 seconds
HR change from 80—40 bpm Average: 12 seconds
Slow alternating ventricular bigeminy 45 bpm
Rapid alternating ventricular bigeminy 163 bpm
Bidirectional systoles 63—73 bpm
Ventricular bigeminy 40—60 bpm
Slow alternating ventricular bigeminy 30 bpm
Rapid alternating ventricular bigeminy 70—163 bpm
Bidirectional systoles 63—73 bpm
30—270 mmHg (4—36 kPa)
10—245 mmHg (1.5—32 kPa)
20—255 mmHg (2.5—34 kPa)
Max. Std. Deviation: 8 mmHg (1.1 kPa)
Max. Mean Error: ±5 mmHg (±0.7 kPa)
40—300 bpm
40—100 bpm: ±5 bpm
101—200 bpm: ±5% of reading
201—300 bpm: ±10% of reading
Typical at HR >60 bpm
Auto/manual: 30 seconds (adult)
Maximum time: 180 seconds (adult)
Typical for normal adult cuff: Less than 10 seconds
165 ±15 mmHg
273