28 Specifications and Standards Compliance
Performance Specifications
Update Rate
Auto Offset Correction
Auto Zero Adjust
Performance Specifications
Maternal Pulse from Toco
Emitted Light Energy
Wavelength Range
Range
Resolution
Display Update Rate
Accuracy
Update Rate
IUP
Performance Specifications
IUP
Measurement Method
Measurement Range
Signal Range
Resolution
Sensitivity
Offset Compensation
Accuracy (not including sensor accuracy)
Update Rate
Auto Offset Correction
ECG
Complies with IEC 60601-2-27:2011 / EN 60601-2-27:2006+AC:2006 except clauses listed below:
201.6.2, 201.8.5.5
201.12.1.101
202.6.2.101
270
Display
Printer
Display
Printer
Display
Printer
1 per second
~4 per second
3 seconds after connecting the transducer, the Toco value is set
to 20 units
Toco value is set to zero following a negative measurement
value for 5 seconds
”15 mW
780—1100 nm
40—240 bpm
1 bpm
1 per second
± 2% or 1 bpm, whichever is greater
every 4 seconds
Passive Resistive Strain Gauge Elements
-100—+300 mmHg
-99—127 mmHg or (-13.2—16.9 kPa)
1 mmHg
1/4 mmHg
5 µV/V/mmHg
+100—-200 mmHg
±0.5% per 100 mmHg
1 per second
~4 per second
3 seconds after connecting the transducer, the IUP value is
set to 0 mmHg