Download  Print this page

Philips FM20/30 Instructions For Use Manual Page 268

Release j.3 with software revision j.3x.xx patient monitoring.
Hide thumbs

Advertisement

28 Specifications and Standards Compliance
Fetal Monitor Performance Specifications
Real time Clock
Buffered Memory
Battery Specifications
Performance Specifications
Avalon FM20/30 Battery
Option #E25
Performance Specification
Avalon CL Transducer
Battery
Fetal / Maternal Specifications
Ultrasound
Complies with IEC 60601-2-37:2007 / EN 60601-2-37:2008
Performance Specifications
Ultrasound
Measurement Method
Measurement Range
Resolution
Jitter @ 200 bpm
Display Update Rate
268
Range
Accuracy
Hold Time
Hold Time
Contents
Operating Time (with new,
fully charged battery)
Charge Time
Operating time
Charge time
Charge time with Avalon FM20/30
Battery Option #E25
US
Display
Printer
from: January 1, 1997, 00:00 to:
December 31, 2080, 23:59
better than ±1 min. per month
infinite if powered by AC; otherwise at least
48 hours
if powered by AC infinite
without power: at least 8 hours
active settings, review alarms, stored trace data
Basic monitoring configuration: >2 hours
(Display Brightness: 70%, Recorder: "On" at 3 cm/min,
NBP: Auto Mode at 15 min, 2 US Transducers, 1 Toco+
with MECG, 1 Patient Module with DECG)
When monitor is off: approx. 6 hours
When monitor is in use: more than 10 hours (depending
on monitor configuration)
With a new and fully charged battery min. 8 hours
From a "low battery" indication to a "fully
charged" indication <3 hours
From a "low battery" indication to a "fully
charged" indication >6 hours.
Ultrasound Pulse Doppler
50—240 bpm
1 bpm
1/4 bpm
”3 bpm
1 per second