Download  Print this page

Performance Specifications - Philips FM20/30 Instructions For Use Manual

Release j.3 with software revision j.3x.xx patient monitoring.
Hide thumbs

Advertisement

Avalon CL Radio
Avalon CL Radio Interface Specifications
Electronic Article Surveillance (EAS)
Short Range Radio Interface* Type
OBR (WMTS)*
OBR (ISM)*
OBR (T108)*
*Wireless transmissions are encrypted for security.

Performance Specifications

Fetal Monitors
Fetal Monitor Performance Specifications
Alarm Signal
Review Alarms
Technology
Frequency Band
Modulation Technique
Effective radiated power
Range
Frequency Band
Effective radiated power
Range
Frequency Band
Effective radiated power
Range
Frequency Band
Effective radiated power
Range
System alarm delay.
The system alarm delay is the processing
time the system needs for any alarm to be
indicated on the fetal monitor, after the
measurement has triggered the alarm.
Pause duration
Extended alarm pause
Sound pressure range
Information
Capacity
28 Specifications and Standards Compliance
EAS tag inside the housing (58 kHz)
Internal SRR interface
IEEE 802.15.4
2.4 GHz ISM (2.400—2.483 GHz)
DSSS (O—QPSK)
max. 0 dBm (1 mW)
ca. 5 m without any physical obstructions as walls
and doors
608—614 MHz
<20 mW (base station)
<1 mW (CL transducers)
min. 100 m/300 ft (in line of sight)
433.05—434.79 MHz
<10 mW (base station)
<1 mW (CL transducers)
min. 100 m/300 ft (in line of sight)
920.6—923.4 MHz
<20 mW (base station)
<10 mW (CL transducers)
min. 100 m/300 ft (in line of sight)
less than 4 seconds
1, 2, 3 minutes or infinite, depending on the
configuration
5 to 10 minutes
min. 0 dB(A) max. 45—85 dB(A)
all alarms/INOPs, main alarms on/off, alarm
silence and time of occurrence
300 items