Product Name
Bacillol® AF
Bacillol®25
Meliseptol®
Accel TB RTU
Oxivir® Tb Cleaner Disinfectant
Oxivir® Tb Wipes
TM/MC
Carpe Diem
Tb
Ready-to-Use General Virucide,
Bactericide, Tuberculocide,
Fungicide, Sanitizer
TM/MC
Carpe Diem
Tb Wipes
Super Sani-Cloth
Germicidal Disposable Wipes
SANI-CLOTH® PLUS
Germicidal Disposable Wipes
SANI-CLOTH® HB Germicidal
Germicidal Disposable Wipes

Cleaning and Disinfecting Monitoring Accessories

To clean, disinfect and sterilize reusable transducers, sensors, cables, leads, and so forth, refer to the
instructions delivered with the accessory.
Do not allow a cleaning or disinfecting agent to leave residues on any of the equipment surfaces. Wipe
residues off, after allowing the appropriate time to for the agent to work, with a cloth.
Cleaning and Disinfecting the Tympanic
Temperature Accessories
Probe and Thermometer Body
1
Wipe the thermometer body clean with a damp cloth. The water temperature should not exceed
55°C (130°F). Do not soak, rinse, or submerge the thermometer under water.
You may add a mild detergent to the water.
Clean the probe tip with a lint free swab. If the probe tip is soiled, clean it with a dampened swab.
2
After you have removed all foreign matter, clean the thermometer lens at the end of the probe tip
3
with a lint free swab or lens wipe. The thermometer lens must be free from fingerprints and/or
smudges for proper operation.
Product Type
Ingredients
liquid, spray
100 g concentrate contains:
Propan-1-ol 45.0 g;
Propan-2-ol 25.0 g; Ethanol 4.7 g.
liquid
Ethanol 100 mg/g
Propan-2-ol (= 2-Propanol) 90 mg/g;
Propan-1-ol (= 1-Propanol) 60 mg/g
spray
50% 1-Propanol
liquid
0.5% accelerated hydrogen peroxide
spray
0.5% accelerated hydrogen peroxide
wipes
0.5% accelerated hydrogen peroxide
spray
0.5% accelerated hydrogen peroxide
wipes
0.5% accelerated hydrogen peroxide
wipes
isopropanol 55%
quaternary ammonium chlorides 0.5%
wipes
isopropanol 15%
quaternary ammonium chlorides 0.25%
wipes
isopropanol < 0.15%
quaternary ammonium chlorides 0.14%