Download  Print this page

Philips FM20/30 Instructions For Use Manual Page 21

Release j.3 with software revision j.3x.xx patient monitoring.
Hide thumbs

Advertisement

With the
start and stop a measurement.
With the
All new SmartKeys are optional, and have to be configured in Configuration Mode for use (see
"SmartKeys" on page 49).
Coincidence INOP Tone
When the cross-channel verification detects that the signal of the maternal heart rate coincides with the
fetal heart rate, the
INOP tone has a configurable delay (see "Cross-Channel Verification (CCV)" on page 135).
Increased Internal Back-up Memory
The internal back-up memory is now able to store traces and data from at least the last 3.5 hours with
the software revision J.3, and minimum 7 hours with the new mainboard hardware revision A 00.18
(see "Manually Recording Stored Data" on page 230 and "Recovering Data" on page 229).
Dual System Interface Support
If the fetal monitor is connected via a LAN connection to OB TraceVue/IntelliSpace Perinatal or
another obstetrical information and surveillance system, the RS232 interface can be used
independently to connect e.g. an EMR system on read-only basis. The system connected to the RS232
interface in this case cannot alter any data (such as ADT data, or the date and time setting), or interfere
with functions of the monitor, but is able to read output data. The obstetrical information and
surveillance system connected via LAN has priority.
USB Interface
An optional USB interface allows the use of bar code readers and input devices such as a keyboard, or
mouse (see "Getting to Know Your Avalon FM20/FM30"/"Bottom" on page 31 and "Getting to
Know Your Avalon FM40/FM50"/"Rear" on page 32).
Flexible Nurse Call Interface
An optional Flexible Nurse Call interface allows the connection of a nurse call device to the fetal
monitors (see "Getting to Know Your Avalon FM20/FM30"/"Bottom" on page 31 and "Getting to
Know Your Avalon FM40/FM50"/"Rear" on page 32).
DHCP Support
The DHCP support offers an alternative to BOOTP. DHCP (dynamic host configuration protocol)
enables the fetal monitors to request an IP address (internet protocol address) from the connected
network (OB TraceVue/IntelliSpace Perinatal) automatically.
Data Export Support
You can now export measurement values from the monitor to other devices via the LAN interface, or
with the optional MIB RS232 interface (see "Getting to Know Your Avalon FM20/FM30"/"Bottom"
on page 31 and "Getting to Know Your Avalon FM40/FM50"/"Rear" on page 32).
SmartKey you can access the NBP Mode selection and setup, and can directly
NBP Modes
SmartKey you can quick admit a patient to the monitor.
QuickAdmit
INOP is now issued with a tone at the fetal monitor. The
Coincidence
2 What's New
Coincidence
21