18 Monitoring Noninvasive Blood Pressure
Problem
Displays zeros for systolic
and diastolic values.
Measurement automatically
repeats
INOP is
NBP Cuff Overpress
displayed
INOP is
NBP Equip Malf
displayed
INOP is
NBP Interrupted
displayed
NBP Measure Failed
200
Possible Causes
Severe vasoconstriction at cuff site
Erratic blood pressure fluctuations due
to arrhythmias or rapid-acting drugs or
contractions
Excessive patient movement or
convulsions
See "Patient Alarms and INOPs" on page 113.
Solutions
Move cuff to another limb, check for shock,
or verify blood pressure using another
method
Try again, if unsuccessful, verify blood
pressure using another method
Wait until contraction has finished
Restrain movement or verify blood pressure
using another method