Starting and Stopping Measurements

Use the Setup menu or the SmartKeys to start and stop measurements.
Action to be performed
Start/Stop manual measurement
Start Auto series
Stop current automatic measurement
Start manual measurement
Start Auto series
Stop manual measurement
Stop current automatic measurement
Stop automatic, or manual measurement AND series
Enabling Automatic Mode and Setting Repetition
Time
1
In the
Toggle between
2
If making an automatic measurement, select
3
set the time interval between two measurements.
NOTE
Be aware that a combination of a recorder speed of less than 3 cm/min and a repetition time of less
than five minutes can result in not all noninvasive blood pressure measurements being recorded on the
fetal trace. For example, if the recorder speed is set to 1 cm/min and the repetition time is set to
two minutes, due to the low speed setting, the recorder will only be able to record every other
noninvasive blood pressure measurement. This affects only the local fetal trace recording, and all
measurements are displayed as normal on the monitor's screen.
Enabling Sequence Mode and Setting Up the
Sequence
1
In the
Select
2
Up to four measurement cycles can be set up which will run consecutively. For each cycle you can
set the number of measurements and the intervals between them. If you want to run less than four
cycles in a sequence, you can set the number of measurements for one or more cycles to
menu, select
Setup NBP
Mode
and
Auto
Manual
menu, select
Setup NBP
Mode
to open the
Setup Sequence
18 Monitoring Noninvasive Blood Pressure
Setup menu
.
, if necessary, to pick the measurement method.
, or press the
Repeat Time
and select
from the pop-up menu.
Sequence
window.
Setup Sequence
SmartKeys
Start/Stop
-
-
Stop All NBP
Repeat Time
Start/ Stop
Start NBP
Stop NBP
Stop All
SmartKey and
.
Off