13 Monitoring Triple FHRs

Troubleshooting

Common problems that may occur when monitoring FHR using ultrasound are listed in "Monitoring
FHR and FMP Using Ultrasound" on page 141.
The following problem may occur when monitoring triplets.
Problem
For more information, see "Additional Information" on page 150.
172
The question mark is
printed repeatedly, and
appears on the screen
and the INOP
is issued.
Coincidence
2
Possible Cause
More than one transducer is
recording the same FHR, or a fetal
transducer records the same heart
rate as the MHR.
trace separation switched off
Classic
here
Solution
Reposition one or more
ultrasound transducer, as
appropriate. See
"Recommended Actions
for Coincidence INOP" on
page 140