Troubleshooting

Common problems that may occur when monitoring FHR using ultrasound are listed in "Monitoring
FHR and FMP Using Ultrasound" on page 141. See also "Monitoring FHR Using DECG" on
page 183 for common problems you might encounter when monitoring FHR directly.
The following problem may occur when monitoring twins.
Problem
The question mark is
printed repeatedly, and
appears on the screen and
the INOP
Coincidence
issued.
For more information, see "Additional Information" on page 150.
Possible Cause
Both transducers are recording the
same FHR, or one fetal transducer is
recording the MHR.
is
12 Monitoring Twin FHRs
Solution
Reposition an ultrasound
transducer. See "Recommended
Actions for Coincidence INOP" on
page 140.