Philips FM20/30 Instructions For Use Manual Page 124

Release j.3 with software revision j.3x.xx patient monitoring.
Hide thumbs

Advertisement

6 Patient Alarms and INOPs
INOP Message
cl SpO Batt Temp
cl SpO Check Batt
cl SpO Disconnect
cl SpO No Cradle
cl SpO Remove
cl SpO Serv Batt
Tympanic Temperature INOPs
INOP Message
<iTemp Label> CalRequired
<iTemp Label> Incompat.
<iTemp Label> Malfunction
<iTemp Label> Meas Failed
<iTemp Label> Overrange
124
Indication
Monitor: INOP tone
CL SpO
Pod: display
2
message
Monitor: INOP tone
CL SpO
Pod: display
2
message
Monitor: INOP tone
CL SpO
Pod: INOP tone
2
and display message
Monitor: INOP tone
CL SpO
Pod: display
2
message
Monitor: INOP tone
CL SpO
Pod: display
2
message
Monitor: INOP tone
CL SpO
Pod: display
2
message
Indication
Monitor: INOP tone
Monitor: INOP tone
Numeric is replaced by a -?-
Monitor: INOP tone
Numeric is replaced by a -?-
Monitor: INOP tone
Numeric is replaced by a -?-
Monitor: INOP tone
Numeric is replaced by a -?-
What to do
The temperature of the battery in the CL SpO
critically high. Check that the Pod is not covered or
exposed to a heat source. If the INOP persists, remove
the CL Pod from the patient, and contact your service
personnel.
The battery in the CL SpO
useful life. Only 50 charge/discharge cycles remain.
Contact your service personnel to replace the battery.
The CL SpO
Pod has lost the SRR connection to the
2
monitor.
The CL SpO
Pod is not in its cradle.
2
The temperature of the battery in the CL SpO
high. Remove the CL SpO
contact service personnel.
The battery in the CL SpO
its useful life. It can no longer be charged. Contact your
service personnel to replace the battery.
What to do
The thermometer requires calibration. Contact your
service personnel.
The thermometer firmware is not supported by the
monitor software.
The thermometer or the interface board is defective.
Check the thermometer on another monitor - if the same
INOP occurs, replace the thermometer. If not, connect
another thermometer to this monitor. If the problem
persists, the interface board has a problem. Contact your
service personnel.
The ambient temperature is out of range. If the ambient
temperature is within the valid range, check the
thermometer. Try picking up a new probe cover. If the
problem persists replace the thermometer and contact
your service personnel.
The measured temperature is out of range. If the
temperature is within the valid range, check with another
thermometer.
Pod is
2
Pod is nearing the end of its
2
Pod is too
2
Pod from the patient and
2
Pod has reached the end of
2