Philips FM20/30 Instructions For Use Manual Page 123

Release j.3 with software revision j.3x.xx patient monitoring.
Hide thumbs

Advertisement

INOP Message
SpO NoisySignal
SpO Poor Signal
SpO Pulse?
SpO Searching
SpO Sensor Malf
SpO Sensor Off
SpO Unkn.Sensor
SpO Upgrade
SpO Interference
CL SpO2 Pod INOPs
INOP Message
cl SpO Batt Empty
!!cl SpO BatEmpty
!!!cl SpO BtEmpty
cl SpO Batt Incmp
cl SpO Batt Low
cl SpO Batt Malf
Indication
Monitor: INOP tone
Numeric is replaced by a -?-
Monitor: display only
Numeric is replaced by a -?-
Monitor: INOP tone
Numeric is replaced by a -?-
Monitor: display only
Numeric unavailable
Monitor: INOP tone
Numeric unavailable
Monitor: INOP tone
Numeric unavailable
Monitor: display only
Numeric is replaced by a -?-
Monitor: display only
Numeric unavailable
Monitor: INOP tone
Numeric is replaced by a -?-
Indication
Monitor: INOP tone
CL SpO
Pod: display
2
message
Monitor: INOP tone
CL SpO
Pod: display
2
message
Monitor: INOP tone
CL SpO
Pod: display
2
message
Monitor: INOP tone
CL SpO
Pod: display
2
message
What to do
Excessive patient movement or electrical interference is
causing irregular pulse patterns. Try to reduce patient
movement, or to relieve the cable strain on the sensor.
The signal condition of the SpO
and measurement accuracy may be compromised.
The detectable pulsations of the SpO
the specified pulse rate range.
SpO
is analyzing the patient signal to derive Pulse, and
2
SpO
values. Please wait until the search analysis is
2
complete.
The SpO
sensor or adapter cable is faulty. Try another
2
adapter cable and sensor. If the INOP persists, contact
your service personnel.
The SpO
sensor is not properly applied to the patient.
2
Apply the sensor following the instructions supplied by
the manufacturer.
The connected sensor or adapter cable is not supported
by the SpO
measurement. Use only specified sensors
2
and cables.
The SpO
measurement is currently in upgrade mode.
2
Monitoring is not possible in this mode.
There is too much interference caused by a high level of
ambient light, and/or electrical interference. Cover the
sensor to minimize ambient light. If the INOP persists,
ensure that the sensor cable is not damaged, or positioned
too close to power cables.
What to do
The remaining battery time of the CL SpO
below 30 minutes. Charge the battery.
The battery in use with the CL SpO
Replace it with one approved for use with the CL SpO
Pod.
The remaining battery time of the CL SpO
2 hours.
There is a malfunction in the CL SpO
system. Contact your service personnel.
6 Patient Alarms and INOPs
measurement is poor
2
signal are outside
2
Pod is
2
Pod is incompatible.
2
Pod is below
2
Pod's battery
2
2
123