Philips FM20/30 Instructions For Use Manual Page 122

Release j.3 with software revision j.3x.xx patient monitoring.
Hide thumbs

Advertisement

6 Patient Alarms and INOPs
INOP Message
cl NBP Batt Temp
cl NBP Check Batt
cl NBP Disconnect
cl NBP No Cradle
cl NBP Remove
cl NBP Serv Batt
SpO2 INOPs
INOP Message
SpO Equip Malf
SpO Erratic
SpO Extd.Update
SpO Low Perf
SpO No Pulse
!! <SpO Label> No Pulse
!!!<SpO Label> No Pulse
SpO No Sensor
122
Indication
Monitor: INOP tone
CL NBP Pod: display
message
Monitor: INOP tone
CL NBP Pod: display
message
Monitor: INOP tone
CL NBP Pod: INOP tone
and display message
Monitor: INOP tone
CL NBP Pod: display
message
Monitor: INOP tone
CL NBP Pod: display
message
Monitor: INOP tone
CL NBP Pod: display
message
Indication
Monitor: INOP tone
Monitor: INOP tone
Numeric is replaced by a -?-
Monitor: display only
Numeric is replaced by a -?-
Monitor: display only
Numeric is replaced by a -?-
Monitor: INOP tone
Numeric is replaced by a -?-
Monitor: INOP tone
Numeric is replaced by a -?-
What to do
The temperature of the battery in the CL NBP Pod is
critically high. Check that the Pod is not covered or
exposed to a heat source. If the INOP persists, remove
the Pod from the patient, and contact your service
personnel.
The battery in the CL NBP Pod is nearing the end of its
useful life. Only 50 charge/discharge cycles remain.
Contact your service personnel to replace the battery.
The CL NBP Pod has lost the SRR connection to the
monitor.
The CL NBP Pod is not in the cradle. You can silence this
INOP, but the INOP message remains visible until the
CL NBP Pod is inserted into the cradle, and the next
measurement is started, or the
selected.
The temperature of the battery in the CL NBP Pod is too
high. Remove the Cableless Measurement Device from
the patient and contact service personnel.
The battery in the CL NBP Pod has reached the end of its
useful life. It can no longer be charged. Contact your
service personnel to replace the battery.
What to do
There is a problem with the SpO
service personnel.
Check the sensor placement. Try another adapter cable
and sensor. If the INOP persists, contact your service
personnel.
The update period of displayed values is extended due to
a noninvasive blood pressure measurement on the same
limb, or an excessively noisy signal.
Accuracy may be compromised due to very low
perfusion. Stimulate circulation at sensor site. If INOP
persists, change the measurement site.
Check the perfusion at measurement site. If necessary,
stimulate circulation or change measurement site. If the
INOP is due to noninvasive blood pressure measurement
on the same limb, wait until the measurement is finished.
Ensure the SpO
sensor is connected. If the INOP
2
persists, try another adapter cable and sensor. If you
silence this INOP, the measurement will be switched off.
SmartKey is
Stop All
hardware. Contact your
2