Release j.3 with software revision j.3x.xx patient monitoring.
Hide thumbs

Advertisement

INOP Message
NBP Equip Malf
NBP Interrupted
NBP Measure Failed
CL NBP Pod INOPs
INOP Message
cl NBP Batt Empty
!!cl NBP BattEmpty
!!!cl NBP BatEmpty
cl NBP Batt Incomp
cl NBP Batt Low
cl NBP Batt Malf
Indication
Monitor: INOP tone
Numeric is replaced by a -?-
Monitor: INOP tone
Numeric is replaced by a -?-
Monitor: INOP tone
Numeric is replaced by a -?-
Indication
Monitor: INOP tone
CL NBP Pod: display
message
Monitor: INOP tone
CL NBP Pod: display
message
Monitor: INOP tone
CL NBP Pod: display
message
Monitor: INOP tone
CL NBP Pod: display
message
What to do
Remove the cuff from the patient. The noninvasive blood
pressure hardware is faulty. Contact your service
personnel.
You can silence this INOP, but the INOP message
remains visible until the next measurement is started, or
the
SmartKey is selected.
Stop All
Check the tubing and cuff for leakages or kinks. Check
that you are using the correct cuff size and placement,
and that the correct patient category is selected. Try
restarting the measurement. If the INOP occurs
repeatedly, contact your service personnel.
You can silence this INOP, but the INOP message
remains visible until the next measurement is started, or
the
SmartKey is selected. This INOP arises when
Stop All
the measurement needed longer than the maximum time
allowed for inflation, deflation, or the total measurement.
Check that you are using the correct cuff size and
placement, and that the correct patient category is
selected. Try restarting the measurement.
You can silence this INOP, but the INOP message
remains visible until the next measurement is started, or
the
SmartKey is selected.
Stop All
Check the condition and suitability of the patient for
noninvasive blood pressure monitoring. Use another cuff
to continue measuring.
What to do
The remaining battery time of the CL NBP Pod is below
30 minutes. Charge the battery.
The battery in use with the CL NBP Pod is incompatible.
Replace it with one approved for use with the CL NBP
Pod.
The remaining battery time of the CL NBP Pod is below
2 hours.
There is a malfunction in the CL NBP Pod's battery
system. Contact your service personnel.
6 Patient Alarms and INOPs
121