6 Patient Alarms and INOPs
INOP Message
DFHR1 Leads Off
!! DFHR1 Leads Off
!!!DFHR1 Leads Off
DFHR2 Leads Off
!! DFHR2 Leads Off
!!!DFHR2 Leads Off
DFHR3 Leads Off
!! DFHR3 Leads Off
!!!DFHR3 Leads Off
DFHR1 Signal Loss
DFHR2 Signal Loss
DFHR3 Signal Loss
DFHR1 Unplugged
DFHR2 Unplugged
DFHR3 Unplugged
MECG Equip Malf
MECG Leads Off
!! MECG Leads Off
!!! MECG Leads Off
MECG Unplugged
NBP INOPs
INOP Message
Cuff Not Deflated
!! Cuff Not Deflat
!!!Cuff Not Deflat
NBP Cuff Overpress
!! Cuff Overpress
!!!Cuff Overpress
120
Indication
Monitor: INOP tone
Numeric is replaced by a -?-
Monitor: INOP tone
Numeric is replaced by a -?-
INOP tone if Alarm Mode is
set to
. No tone if Alarm
All
Mode is set to
INOP only
Monitor: INOP tone
Numeric is replaced by a -?-
Monitor: INOP tone
Monitor: INOP tone
Numeric is replaced by a -?-
Monitor: INOP tone
Numeric is replaced by a -?-
Indication
Monitor: INOP tone (severe
yellow/red INOP tone)
Numeric is replaced by a -?-
During this INOP, alarms
cannot be paused or switched
off.
Monitor: INOP tone (severe
yellow/red INOP tone)
Numeric is replaced by a -?-
During this INOP, alarms
cannot be paused or switched
off.
What to do
One or more DECG lead is not attached. Make sure that
all required leads are attached, and no electrodes have
been displaced. Check all connections are sound, and that
the leg plate attachment electrode is properly attached. If
the INOP persists, try using another adapter cable, or leg
plate attachment electrode. If the INOP persists, contact
your service personnel.
The input signal quality is not sufficient to process the
measurement. Reapply the fetal scalp electrode.
.
Reconnect the DECG transducer to the monitor. Check
if all connections are sound.
There is a problem with the MECG hardware. Contact
your service personnel.
One or more MECG lead is not attached. Ensure that all
required leads are attached, and no electrodes have been
displaced. Check if all connections are sound. If the
INOP persists, try using another adapter cable. If the
INOP still persists, contact your service personnel.
Reconnect the MECG transducer to the monitor. Check
if all connections are sound.
What to do
Remove the cuff from the patient. Make sure that the
tubing is not kinked or twisted. Try restarting the
measurement. You can silence the INOP, but the INOP
message remains visible until the next measurement is
started, or the
Stop All
The cuff pressure exceeds the overpressure safety limits.
Remove the cuff from the patient. Ensure that the tubing
is not kinked or twisted, and that the correct patient
category is selected.
Try restarting the measurement. You can silence this
INOP, but the INOP message remains visible until the
next measurement is started, or the
selected.
SmartKey is selected.
SmartKey is
Stop All