Cableless Transducer INOPs
INOP Message
cl US Batt Empty
!!cl US Batt Empty
!!!cl US BattEmpty
cl Toco Batt Empty
!!clToco BattEmpty
!!!clTocoBattEmpty
cl ECG/IUP Empty
!!cl ECG/IUP Empty
!!!clECG/IUP Empty
cl US Batt Low
cl Toco Batt Low
cl ECG/IUP BattLow
cl US Batt Malf
cl Toco Batt Malf
clECG/IUP BattMalf
cl US Check Temp
cl Toco Chk Temp
clECG/IUP Chk Temp
cl US Disconnect
cl Toco Disconnect
cl ECG/IUP Disconn
cl US Malfunction
cl Toco Malf
cl ECG/IUP Malf
DECG and MECG INOPs
INOP Message
DFHR1 Equip Malf
DFHR2 Equip Malf
DFHR3 Equip Malf
Indication
Monitor: Display only
CL transducer: LED flashes
red
Monitor: Display only
CL transducer: LED flashes
red
Monitor: INOP tone
CL transducer: tone and cyan
LED
Monitor: INOP tone
CL transducer: tone and cyan
LED
Monitor: INOP tone
CL transducer: tone and cyan
LED
Monitor: INOP tone
Numeric is replaced by a -?-
CL transducer: tone and cyan
LED
Indication
Monitor: INOP tone
What to do
The remaining monitoring time with this transducer is
below 15 minutes. Charge battery.
The remaining monitoring time with this transducer is
below 30 minutes. Charge battery.
Malfunction of the battery system (charger circuit or
battery) detected (e.g. persistent communication error,
over-voltage, over-current, battery incompatible). Contact
your service personnel.
The temperature of the transducer or its battery is
critically high. If the transducer is docked on the base
station, charging is interrupted. Check that the transducer
is not covered and do not expose it to a heat source. Do
not place the base station in a sunny window in a room
without active air conditioning. The environment
temperature to charge a transducer battery should not
exceed 40°C/104°F. If the INOP persists, remove the
transducer from patient and contact your service
personnel.
The cableless transducer has lost the connection to the
monitor.
Malfunction of the transducer. Contact your service
personnel.
What to do
There is a problem with the DECG hardware. Contact
your service personnel.
6 Patient Alarms and INOPs
119