6 Patient Alarms and INOPs
INOP Message
Check Touch Input
Coincidence
!! Coincidence
Internal.Comm.Malf
NST Elapsed
Paper End
Settings Malfunct
Speaker Malfunct
SRR Incompatible
SRR Interference
SRR Invalid Chan
SRR Malfunction
Tele Incompatible
Tele Malfunction
Unsupported LAN
User I/F Malfunct
116
Indication
Monitor: Display only
Monitor: INOP tone
Monitor: INOP tone
Monitor: INOP tone
(Message and tone only if
,
Setup NST Timer
Notification
is set to
in
Alarm
Configuration Mode).
Monitor: INOP tone
Monitor: INOP tone
Monitor: INOP tone
Monitor: INOP tone
Monitor: INOP tone
Monitor: INOP tone
Monitor: INOP tone
Monitor: INOP tone
Monitor: INOP tone
Monitor: INOP tone
Monitor: INOP tone
What to do
Perform a visual and functional check of the touch input
device. Contact your service personnel.
The fetal heart rate(s) and maternal heart rate/pulse
persistently coincide with another. (see "Cross-Channel
Verification (CCV)" on page 135). The INOP tone has a
configurable delay. The default setting is an instant INOP
tone.
There is a problem with I2C Bus communication in the
monitor. Contact your service personnel.
The time has expired for the NST timer. Clearing the
timer clears the INOP.
The end of the paper pack is detected. Insert a new pack
of paper.
The monitor cannot use the predefined settings for
monitoring. Contact your service personnel.
Contact your service personnel to check the speaker and
the connection to the speaker.
The connected SRR (Short Range Radio) board or
firmware cannot be used with this base station's software
revision. Contact your service personnel.
The short range radio connection has interference from
another device. Try using another channel.
The channel configuration of the Short Range Radio is
invalid. Check channel and channel mask configuration.
Malfunction in the SRR (Short Range Radio) module(s).
If the INOP persists, contact your service personnel.
The base station is not supported by the monitor. Check
software revisions.
Malfunction of the connected base station. If the INOP
persists contact your service personnel.
There is a problem with the communication to the
network and central monitoring is currently not possible.
Check the connection. If the INOP persists, switch off
the monitor and contact your service personnel.
Perform a visual and functional check of all the monitor
input devices. Contact your service personnel.