Patient Alarms and INOPs

6
This chapter lists alarms and technical alarms (INOPs) for the fetal monitors irrespective of their
priority.

Alarm Messages

Fetal alarms are identified by either "FHR" or "DFHR". All other alarms without these identifiers refer
to maternal parameters.
Alarm Message
*** Brady (Pulse)
***xBrady xxx<yyy
*** Desat
*** Desat xx < yy
** DFHR1 High
** DFHR2 High
** DFHR3 High
**DFHR1 xxx>yyy
**DFHR2 xxx>yyy
**DFHR3 xxx>yyy
** DFHR1 Low
** DFHR2 Low
** DFHR3 Low
**DFHR1 xxx<yyy
**DFHR2 xxx<yyy
**DFHR3 xxx<yyy
*** Extreme Brady
***xBrady xxx<yyy
From
Condition
SpO
The heart rate from the Pulse signal has
2
fallen below the bradycardia limit. xxx
denotes the lowest measured value; yyy is
the bradycardia limit.
SpO
The SpO
2
desaturation alarm limit. xx denotes the
lowest measured value, and yy is the
desaturation limit.
FHR (DECG)
The fetal heart rate obtained from DECG
has risen above the high alarm limit.
xxx denotes the highest measured value, and
yyy is the high alarm limit.
FHR (DECG)
The fetal heart rate obtained from DECG
has fallen below the low alarm limit.
xxx denotes the lowest measured value, and
yyy is the low alarm limit.
MECG
The maternal heart rate obtained from the
maternal ECG has fallen below the extreme
bradycardia limit.
xxx denotes the lowest measured value, and
yyy is the extreme bradycardia limit.
value has fallen below the
2
Indication
Numeric flashes and alarm
limit is highlighted, red alarm
message, alarm tone.
Numeric flashes, red alarm
message, alarm tone.
Numeric flashes and high
limit is highlighted, yellow
alarm message, alarm tone.
Numeric flashes and low limit
is highlighted, yellow alarm
message, alarm tone.
Numeric flashes, red alarm
message, alarm tone.