4 INSTALLATION
Caution: Before installation, ensure that the specified instrument supply voltage matches the
4.1 MECHANICAL INSTALLATION
Figure 1 gives installation details.
4.1.1 Installation Procedure
1.
If it is not already in place, fit the IP65 sealing gasket behind the front bezel of the instrument.
2.
Insert the instrument through the panel cutout, from the front of the panel.
3.
Spring the retaining clips into place, and secure the instrument by holding it firmly in place whilst pushing
both clips towards the rear face of the panel.
4.
The protective membrane can now be removed from the display.
For the sake of clarity, the
panel is shown as though
transparent
Push springs towards panel
4.1.2 Demounting
Warning: Before removing the supply voltage wiring, isolate the supply voltage and secure it
1.
Isolate the mains supply and secure it against accidental operation. Remove all wiring and the USB
device and Ethernet cable (if any).
2.
Remove the retaining springs by unhooking them from the sides using a small flat-blade screwdriver.
3.
Pull the instrument forwards out of the panel.
A
September 18
FEC920: USER GUIDE
facility supply.
Figure 1 Securing the Instrument
against unintended operation.
A5E45696052A Rev-AA
Retaining spring (two places)
Push springs towards panel