Checking Alignment For Beam Strength - HP 5528A User Manual

Laser measurement system
Table of Contents

Advertisement

Model
5528A
Dista
nce Measureme nts
OISTANCE INTERFEROMETER WITH TARGET
NOTE: THE TARGET SHOULD BE
POSIT
IONED WITH
THE 3MM HOLE UPPERMOST ON THE
INTERFERO-
METER AND WITH THE CROSSHAIRS UPPERMOST
ON THE RETROREFLECTOR
.
DISTANCE RETROREFLECTOR WITH TAAGET
Figure
6-7.
Inferferomeler
and
Retroreflec!or,
wifh
largels
Checking Alignment for Beam Strength
1.
You
are now ready to make
sure you
have
sufficient
beam
strength over
the full length
of
travel.
2.
Open both
laser Head ports
to
their
largest
openings.
3.
Move the
Optics over
their full travel length
while
watching the BEAM
STRENGTH
meter.
Any reading
in the
meter's green area
indicates
sufficient
beam
strength.
4.
If the
meter's reading goes
into the red area
al
any time, return the
Oplics
to their
"nea r-end-
of-travel" position and reposition
them
so
their return dots overlap
on
the
laser
He .. d's
crosshair
target
(after you rolale
the
target
into
posi tion
at
the return
port and rotate
the
sma
ll
aperture
into
position at the exit pori
).
Then
repea
t
both alignment
checks.
13